Abdomen – Siemens syngo CT 2005C User Manual

Page 119

Advertising
background image

Abdomen

119

* CARE Bolus may be used to optimize the bolus tim-
ing. Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.

Non
Contrast

Arterial

Phase

Venous
Phase

kV

130

130

130

Effective mAs

80

80

80

Rotation time

1.0 sec

0.8/1.0
sec

1.0 sec

Slice

collimation

5.0 mm

5.0 mm

5.0 mm

Slice width

8.0 mm

6.0 mm

6.0 mm

Feed/Rotation

15.0 mm

14.0/

15.0 mm

15.0 mm

Pitch Factor

1.5

1.4/1.5

1.5

Increment

8.0 mm

6.0 mm

6.0 mm

Kernel

B41s

B41s

B41s

CTDI

Vol

8.62 mGy

8.62 mGy

8.62 mGy

Effective dose

Male:

3.34 mSv

Female:

4.19 mSv

Male:

3.34 mSv

Female:

4.19 mSv

Male:

3.34 mSv

Female:

4.19 mSv

Contrast medium IV injection

Start delay

20 – 25 sec.* (arterial phase)

50 – 75 sec.* (venous phase)

Flow rate

3.0 ml/sec.

Total amount

100 ml