Vascular – Siemens syngo CT 2005C User Manual

Page 173

Advertising
background image

Vascular

173

Hints

• CARE Bolus may be used to optimize the bolus tim-

ing.

• Set the ROI for monitoring scan in the abdominal

aorta with triggering threshold of 120 HU, or use
manual triggering.

• Do not administer oral contrast medium, as this

impairs the editing of MIP/SSD/VRT images.

• Use water as oral contrast.

• The length of the CM spiral in the topogram (via the

table position) can be planned exactly by means of
pre-contrast images.

• Pre-contrast images are used to visualize calcifica-

tion.

• Excellent post-processed images can be created

using a thin slice thickness and overlapping images,
i.e. the increment should be smaller than the slice
thickness.

Contrast medium IV injection
Start delay

20 – 25 sec.

Flow rate

3.0 – 3.5 ml/sec.

Total amount

100 – 120 ml

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 173 Friday, April 8, 2005 9:55 AM

Advertising