Abdomen – Siemens syngo CT 2005C User Manual

Page 115

Advertising
background image

Abdomen

115

Hints

• You could repeat the same protocol by simply click-

ing the chronicle with the right mouse button for
repeat“, e.g., when both non-contrast and contrast
studies are required.

• Delayed scans may be required for the kidneys &

bladder.

AbdRoutine

kV

130

Effective mAs

60

Rotation time

0.8/1.0 sec

Slice collimation

4.0 mm

Slice width

5.0 mm

Feed/Rotation

14.4 mm

Pitch Factor

1.8

Increment

5.0 mm

Kernel

B41s

CTDI

Vol

6.47 mGy

Effective dose

Male: 2.61 mSv

Female: 3.32 mSv

Contrast medium IV injection

Start delay

50 – 60 sec.

Flow rate

1.0 – 2.0 ml/sec.

Total amount

100 ml

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 115 Friday, April 8, 2005 9:55 AM