Lower Extremities

FootHR

Indications:

Spiral mode for high resolution studies of the foot,
e.g., masses, trauma, disorders of the joint etc.

A range of 10 cm will be covered in 52 sec.

Hint

• For image reconstruction of soft tissue, use kernel

B31s.

FootHR

2

nd

reconstruction

kV

130

Effective mAs

30

Rotation time

1.0 sec

Slice collimation

1.0 mm

Slice width

3.0 mm

1.25 mm

Feed/Rotation

2.0 mm

Pitch Factor

1.0

Increment

3.0 mm

0.8 mm

Kernel

U90s

U90s

CTDI

Vol

3.23 mGy

Effective dose

Male: 0.01 mSv

Female: 0.01 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 160 Friday, April 8, 2005 9:55 AM