Dose information – Siemens syngo CT 2005C User Manual

Page 34

Advertising
background image

34

Dose Information

• For the average patients examination, CARE Dose

does not require any manual changes to the scan
protocol. However, the mAs must be adapted manu-
ally for obese and pediatric patients.

• The mean value of the mAs applied will be lower

than what you have selected. Although the average
mA for the entire scan will be lower than selected,
we allow the scanner to apply increased mA levels
for the high attenuation views. This may cause dif-
ferent results of the tube load controller when
switching on and off CARE Dose.

• The mean value of the effective mAs applied is

shown in the image text.

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 34 Friday, April 8, 2005 9:55 AM