syngo Dental CT

• It is mandatory to position the patient head in the

center of the scan field – use the lateral laser light
marker for positioning.

• Gantry tilt is not necessary since you have the ability

to tilt the reference line to generate an axial refor-
matted image at the desired plane. However, in
order to minimize the scan length for the same ana-
tomical region, it is recommended to position the
patient’s head at the appropriate scan plane when-
ever possible:
– For the upper and lower jaw:

occlusal plane in parallel to the scan plane.

– For either jaw:

jaw bone in parallel to the scan plane.

• It is recommended to end the exam first, and then

start the Dental evaluation.

User interface of syngo Dental CT