Head – Siemens syngo CT 2005C User Manual

Page 75

Advertising
background image

Head

75

Hints

• An automatic bone correction allows for improved

head image quality, without any additional post-pro-
cessing.

• In order to optimize image quality versus radiation

dose, scans are provided within a maximum scan
field of 300 mm with respect to the iso-center. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.

BaseSeq

CerebrumSeq

kV

130

130

Effective mAs

240

240

Rotation time

1.5 sec

1.5 sec

Slice collimation

1.5 mm

4.0 mm

Slice width

3.0 mm

8.0 mm

Feed/Scan

3.0 mm

8.5 mm

Kernel

H31s

H31s

CTDI

Vol

54.65 mGy

51.43 mGy

Effective dose

Male:
0.77 mSv

Female:
0.80 mSv

Male: 1.52 mSv

Female:
1.68 mSv

Contrast medium IV injection

Start delay

60 sec.

Flow rate

2 ml/sec.

Volume

50 – 60 ml