Test bolus using care bolus, Contrast medium test bolus using care bolus – Siemens EMOTION 16-Jun User Manual

Page 128

Advertising
background image

128

Contrast Medium

Test Bolus using CARE Bolus

You can use the CARE Bolus option as a Test Bolus.

Method

1.Insert a Bolus Tracking via the context menu prior to

the spiral.

2.Insert contrast from the context menu.

Note: By inserting contrast you are interrupting the
Auto range function preventing an automatic start
of the spiral.

3.Start with the topogram.

4.Position the premonitoring scan and the spiral.

5.Perform the premonitoring scan, position and

accept the ROI.

6.Start the monitoring scans and a small amount of

contrast medium (20 ml/2.5 ml/sec.). The rate of the
injection on the monitoring scans should match the
rate of injection used for the spiral scan.
Note: When you start the spiral manually, the sys-
tem switches to the Trigger tab card. The trigger line
is not shown at this stage.

7.Now you can read the proper delay from the Trigger

tab card.

8.Insert the delay in the Routine tab card and load the

spiral.

9.Start the spiral and injector at the same time with

the appropriate amount of contrast for the study
being performed.

Advertising