Extrroutinehr, Extrroutinehr 568, Children – Siemens EMOTION 16-Jun User Manual

Page 568

Advertising
background image

568

Children

ExtrRoutineHR

Indications:

Spiral mode for high-resolution bone studies, for
example, tumors, post-traumatic changes, orthopedic
indications, etc.

For SOMATOM Emotion 16-slice configuration:

A typical range of 6 cm covered in 35.33 sec.

For SOMATOM Emotion 6-slice configuration:

A typical range of 10 cm covered in 35.33 sec.

* The conversion factor for a 7-year-old child, and a

scan range of 60 mm was used.

** Adjust the mAs value to the body region.

Emotion 16

ExtrHR

kV

110

Effective mAs/
Quality ref. mAs**

50

Rotation time

1.0 sec.

Acquisition

4 x 0.6 mm

Slice collimation

0.6 mm

Slice width

0.6 mm

Feed/Rotation

1.8 mm

Pitch Factor

0.75

Increment

0.4 mm

Kernel

U90s

CTDI

Vol

4.70 mGy

Effective dose

Male: 0.01 mSv*
Female: 0.01 mSv*

Advertising