Children – Siemens EMOTION 16-Jun User Manual

Page 563

Advertising
background image

Children

563

* The conversion factor for a 7-year-old child, and a

scan range of 160 mm was used.

Emotion 16

Spine

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

60

Rotation time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

2.0 mm

1.0 mm

Feed/Rotation

6.2 mm

Pitch Factor

0.65

Increment

2.0 mm

0.7 mm

Kernel

B41s

B60s

CTDI

Vol

4.80 mGy

Effective dose

Male: 1.28 mSv*
Female: 1.43 mSv*

Emotion 6

SpineRoutine

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

78

Rotation time

1.0 sec.

Acquisition

6 x 2.0 mm

Slice collimation

2.0 mm

Slice width

3.0 mm

2.5 mm

Feed/Rotation

10.2 mm

Pitch Factor

0.85

Increment

3.0 mm

1.5 mm

Kernel

B41s

B60s

CTDI

Vol

5.54 mGy

Effective dose

Male: 2.36 mSv*
Female: 4.86 mSv*