Dose Information

For spiral scan protocols, the indicated mAs is the
effective mAs per image. The correlation between tube
current and effective mAs of spiral scans on a Multi-
slice CT scanner is expressed by the following formula:

where Slice collimation refers to the collimation of one
detector row, and nrow is the number of used detector
rows.

Effective mAs = mA x RotTime

Pitch Factor

Pitch Factor = Feed per Rotation
nrow x Slice collimation

mA = effective mAs x Pitch Factor
RotTime