Workflow Information

75

Oblique reconstruction of the sinuses:

– Scan a topogram
– Plan your axial spiral scan range
– Reconstruction of the spiral images

(RT images)

– Select Recon job Type oblique
– Select the sagittal image segment
– Enable Free Mode
– Rotate the reference lines until the best view of the

sinuses is displayed in one of the other segments

– Select this segment and press the Set FoV Seg-

ment button

– Adjust the FoV to your needs
– Define your desired reconstruction parameters

(e.g., image type SPO)

– Start reconstruction

Oblique reconstruction of the vascular tree:

– Scan a topogram
– Plan your spiral scan range
– Axial reconstruction of the spiral images

(RTD images)

– Select Recon job Type oblique
– Select button MIP Thin as image type for the

planning volume on the toolbar

– Enable Free Mode
– Rotate the reference lines until the best view

of the entire vascular tree is displayed in one of
the other segments

– Select the coronal image segment
– Select this segment and press the Set FoV Seg-

ment button

– Adjust the FoV to your needs
– Define your desired reconstruction parameters

(e.g., Type MIP Thin)

– Start reconstruction