Special modes of care dose 4d, Dose information – Siemens EMOTION 16-Jun User Manual

Page 53

Advertising
background image

Dose Information

53

Special Modes of CARE Dose 4D

For certain examination protocols CARE Dose 4D uses
modified tube current modulation, to meet specific
conditions, for example:

• for Adult Head protocols the tube current is adapted

to the variation along the patient’s long axis and not
to the angular attenuation profile.

• for Extremities, CARE Vision, syngo Neuro Perfusion

CT, syngo Body Perfusion CT and other special proto-
cols (indicated as CARE Dose), only angular tube cur-
rent modulation is supported.

• for Osteo and Cardio protocols the mAs setting is

adjusted to the patient size and not modulated dur-
ing the scan, except if ECG pulsing is switched on.