Carevisionbone, Carevisionbone 442, Specials – Siemens EMOTION 16-Jun User Manual

Page 442

Advertising
background image

442

Specials

CAREVisionBone

The images will be reconstructed and displayed in the
CARE View mode with three times 4.8/6.0 mm or one
time 10 mm slice thickness and a kernel of B50.

Emotion 16

CAREVision

kV

130

Effective mAs/
Quality ref. mAs

30

Rotation time

0.6 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

4.8 mm

Feed/Scan

0.0

Increment

1.0 mm

Kernel

B50s

CTDI

vol

3.75 mGy

Emotion 6

CAREVision

kV

130

Effective mAs/
Quality ref. mAs

30

Rotation time

0.6 sec.

Acquisition

6 x 3.0 mm

Slice collimation

3.0 mm

Slice width

6.0 mm

Feed/Scan

0.0

Increment

1.0 mm

Kernel

B50s

CTDI

vol

3.06 mGy

Advertising