Dose information – Siemens EMOTION 16-Jun User Manual

Page 50

Advertising
background image

50

Dose Information

2) Angular tube current modulation:
Based on the above described axial eff. mAs profiles,
the tube current is modulated during each tube rota-
tion. Therefore the angular attenuation profile is mea-
sured automatically during the scan and the tube cur-
rent is modulated accordingly in real time to achieve
an optimum distribution of the X-ray intensity for every
viewing angle.

Relation between relative attenuation and relative
tube current. The adaptation strength may be
adjusted by user separately for the left branch (slim)
and the right branch (obese) of the curve. This adjust-
ment effects all examinations. The gray lines here
indicates the theoretical limits of the adaptation (con-
stant dose resp. constant image noise). The absolute
(eff.) mAs value is scaled with the
Image Reference
mAs
value, which may be adjusted in the Scan Card by
the user.

re

l.

tube curr

en

t

reference attention

rel. attenuation

Im

ag
e Qua

lity

re

fe

re

nce tube curr

ent

constant dose

o

b

e

se

co

n

st

an

t

im

ag

e

n

o

is

e

slim

stro

ng i

ncre

ase

avera

ge inc

rease

Weak i

ncreas

e

we

ak d

ecre

ase

av

er

ag

e

de

cr

ea

se

st

ro

ng

d

ec

re

as

e