Additional important information, Radiation therapy – Siemens EMOTION 16-Jun User Manual

Page 482

Advertising
background image

482

Radiation Therapy

Additional Important Information

HU values

The HU values are of crucial importance for the therapy
planning systems of radiation therapists!

• With huge objects, the CT value is independent of

the kernel. With smaller objects, edge effects pro-
duced by the kernel influence the HU values as well
as the scanning. Feed and collimation do not have
any influence.

• An extended FoV of 800 mm means that only 500

mm are scanned, the rest will be interpolated.

• The effect on the HU values of having carbon plate

tagged additionally to the patient table should be
negligible and not measurable.

• CARE Dose 4D does not have any effect on the HU

values.

• HU values and the conversion of the electron densi-

ties depend on the applied spectrum, e.g. kV, prefil-
tration(kV, prefiltration, etc.)

• HU values depend on the use of the online bone cor-

rection (PFO).