Specials – Siemens EMOTION 16-Jun User Manual

Page 448

Advertising
background image

448

Specials

Radiation exposure to patients

• CARE Vision applies continuous exposure at moder-

ate mA levels.

• Due to the potentially long scan times and the lim-

ited scan volume, the dose for certain slices may
increase to levels significantly higher than those
known from standard CT applications.

• The patient’s exposure levels is usually estimated by

CTDIvol.

• This unit is designed to give the average dose in the

scanned volume.

• Before starting the scan, the dose rate (CTDIvol in

mGy per second) is displayed on the monitor.

• During the CARE Vision scan, the accumulated dose

(CTDI vol) reflecting the patient exposure is dis-
played on the monitor. The display scale ranges from
zero to 3000 mGy.

• If the table is shifted during the examination, the

accumulated dose will be distributed to different
slices and will be lower than indicated by the display.

• When a new scan is loaded, the dose display starts

again from zero.

Advertising