Spine

Osteo

This is the scan protocol for the syngo Osteo CT appli-
cation package to assist the physician with the quanti-
tative assessment of vertebral bone mineral density
(BMD) in the diagnosis and follow-up of osteopenia
and osteoporosis.

Emotion 16

Osteo

kV

80

Effective mAs/
Quality ref. mAs

220

Rotation time

1.0 sec.

Acquisition

2 x 5.0 mm

Slice collimation

5.0 mm

Slice width

10.0 mm

Feed/Scan

0.0 mm

Kernel

S80s

CTDI

Vol

5.72 mGy

Effective dose

Male: 0.48 mSv

Female: 0.71 mSv

Emotion 6

Osteo

kV

80

Effective mAs/
Quality ref. mAs

220

Rotation time

1.0 sec.

Acquisition

2 x 5.0 mm

Slice collimation

5.0 mm

Slice width

10.0 mm

Feed/Scan

0.0 mm

Kernel

S80s

CTDI

Vol

5.50 mGy

Effective dose

Male: 0.21 mSv
Female: 0.49 mSv