Vascular – Siemens EMOTION 16-Jun User Manual

Page 371

Advertising
background image

Vascular

371

Hint

• Use of CARE Bolus with monitoring scans positioned

at the level of the basilar artery or carotid artery. Set
the trigger threshold at 120 HU, or use manual trig-
gering.

Emotion 6

HeadAngio

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

70

Rotation time

0.8 sec.

Acquisition

6 x 1.0 mm

Slice collimation

1.0 mm

Slice width

3.0 mm

4.0 mm

Feed/Rotation

5.1 mm

Pitch Factor

0.85

Increment

3.0 mm

4.0 mm

Kernel

H31s

H20s

CTDI

Vol

12.88 mGy

Effective dose

Male: 0.37 mSv
Female: 0.42 mSv

Contrast medium IV injection
Start delay

18 sec.

Flow rate

3.5 ml/sec.

Total amount

75 ml