566

Children

SpineThinSlice

Indications:

Spiral mode for the spine when Multi Planar Reforma-
tion (MPR) are intended, for example, post-traumatic
changes, tumors, malformations, etc.

For SOMATOM Emotion 6-slice configuration:

A typical range of 15 cm covered in 47.45 sec.

* The conversion factor for a 7-year-old child, and a

scan range of 160 mm was used.

Emotion 6

SpineThinSlice

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

78

Rotation time

1.0 sec.

Acquisition

6 x 1.0 mm

Slice collimation

1.0 mm

Slice width

2.0 mm

1.25 mm

Feed/Rotation

3.3 mm

Pitch Factor

0.55

Increment

2.0 mm

0.8 mm

Kernel

B41s

B60s

CTDI

Vol

6.47 mGy

Effective dose

Male: 2.55 mSv*
Female: 3.02 mSv*