Abdomen – Siemens EMOTION 16-Jun User Manual

Page 285

Advertising
background image

Abdomen

285

A range of 20 cm including liver, pancreas and kidney:
arterial phase acquired in 12.96 sec.

* CARE Bolus may be used to optimize the bolus tim-

ing. Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trig-
gering threshold of 120 HU, or use manual trigger-
ing.

Emotion 6

Venous Phase

kV

130

Effective mAs/
Quality ref. mAs

120

Rotation Time

0.8/0.6 sec.

Acquisition

6 x 2.0 mm

Slice collimation

2.0 mm

Slice width

5.0 mm

Feed/Rotation

10.2 mm

Pitch Factor

0.85

Increment

5.0 mm

Kernel

B41s

CTDI

Vol

13.08 mGy

Effective dose

Male: 5.06 mSv
Female: 6.30 mSv

Contrast medium IV injection

Start delay

20 – 25 sec.* (arterial phase)

50 – 75 sec.* (venous phase)

Flow rate

4 – 5 ml/sec.

Total amount

100 –120 ml