Children

Hints

• Children, who are older than age 6, should be

scanned with an adult protocol as the skull by this
time is fully grown.

• When bone structure is of interest, use kernel C60s

for image reconstruction.

• An advanced algorithm allows for improved head

image quality, without additional post-processing.

• In order to optimize image quality versus radiation

dose, scans are provided within a maximum scan
field of 300 mm with respect to the iso-center. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.

• To work without CARE Dose 4D use for children

< 6 month 90 mAs
6 month-3 years 150 mAs
3-6 years 220 mAs.

Contrast medium IV injection
Start delay

exam dependent

Flow rate

dependent upon needle size/

Access site

Total amount

1 – 2 ml per kg of body weight