Body kernels, Thorax – Siemens EMOTION 16-Jun User Manual

Page 231

Advertising
background image

Thorax

231

Body Kernels

• As standard kernels for body tissue studies B30s or

B40s are recommended; softer images are obtained
with B20s or B10s (extremely soft). The kernels B31s
or B41s have about the same visual sharpness as
B30s, respectively, B40s, the image appearance,
however, is more agreeable due to a ”fine-grained”
noise structure; quite often, the low contrast detect-
ability is improved by using B31s, B41s instead of
B30s, B40s.

• For higher sharpness, as is required e.g. in patient

protocols for cervical spine, shoulder, extremities,
thorax, lung, the kernels B50s, B60s, B70s, B80s are
available.

• A special High Resolution lung kernel is supplied

with B90s; by using kernel B90s, practically the same
image impression is obtained as with HCE-filtering of
a B40s image.

Advertising