Hint

• For low dose studies use only 20 mAs.

Emotion 6

Orbit

2

nd

reconstr.

kV

130

Effective mAs/
Quality ref. mAs

70

Rotation Time

1.0 sec.

Acquisition

6 x 1.0 mm

Slice collimation

1.0 mm

Slice width

5.0 mm

1.25 mm

Feed/Rotation

5.1 mm

Pitch Factor

0.85

Increment

5.0 mm

0.5 mm

Kernel

H70s

H70s

CTDI

Vol

19.25 mGy

Effective dose

Male: 0.25 mSv
Female: 0.32 mSv

Contrast medium IV injection
Start delay

45 sec.

Flow rate

2.0 ml/sec.

Total amount

75 ml