Children – Siemens EMOTION 16-Jun User Manual

Page 555

Advertising
background image

Children

555

* The conversion factor for a 7-year-old child, and a

scan range of 200 mm was used.

Emotion 16

ThoraxHRSeq

kV

110

Effective mAs/
Quality ref. mAs

36

Rotation time

1.0 sec.

Acquisition

4 x 0.6 mm

Slice collimation

0.6 mm

Slice width

1.2 mm

Feed/Scan

10.0 mm

Kernel

B90s

CTDI

Vol

0.81 mGy

Effective dose

Male: 0.31 mSv*
Female: 0.39 mSv*

Emotion 6

ThoraxHRSeq

kV

110

Effective mAs/
Quality ref. mAs

36

Rotation time

1.0 sec.

Acquisition

1 x 1.0 mm

Slice collimation

1.0 mm

Slice width

1.0 mm

Feed/Scan

10.0 mm

Kernel

B90s

CTDI

Vol

0.24 mGy

Effective dose

Male: 0.20 mSv*
Female: 0.24 mSv*