198

Head

Orbit

Indications:

Spiral mode for orbital studies, e.g. fracture.

For SOMATOM Emotion 16-slice configuration:

A range of 50 mm will be covered in 8.51 sec.

For SOMATOM Emotion 6-slice configuration:

A range of 50 mm will be covered in 11.80 sec.

Emotion 16

Orbit

2

nd

reconstr.

kV

130

Effective mAs/
Quality ref. mAs

70

Rotation Time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

1.0 mm

Feed/Rotation

7.7 mm

Pitch Factor

0.80

Increment

5.0 mm

0.7 mm

Kernel

H70s

H70s

CTDI

Vol

18.62 mGy

Effective dose

Male: 0.50 mSv
Female: 0.54 mSv