6712SR Refrigerated Sampler
Section 5 Extended Programming

5-14

4

“LEVEL” SET POINT:

1.000 ft

( 0.001 - 30.00)

Enter a set point of 1 foot.

5

ENABLED WHEN:

ABOVE SET POINT
BELOW SET POINT

Select ABOVE SET POINT.

6

EVENT02:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

Select LEVEL again.

7

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

Choose RATE OF CHANGE.

8

CONDITION IS TRUE

WHEN “LEVEL”

RISES FALLS

Choose RISES.
This will enable the event condition if a rise occurs within the time frame
and above the amount specified in screen #9.

9

“LEVEL” RISES

1.000 ft

__HOURS, __MINUTES

Enter 1 ft and press

.

Then enter 2 hours.

10

EVENT03:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

Select pH.

11

“pH5” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

Select RANGE.

12

“pHx” RANGE:

LOWER: 4.00
UPPER: 10.00

( 0.00 - 14.00)

Enter 4 for the lower range and 10 for the upper range.

13

ENABLED WHEN:

INSIDE RANGE

OUTSIDE RANGE

Choose INSIDE RANGE.

14

EVENT04:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 DONE

You will not be entering a fourth event condition, so use the arrow keys to
scroll down, and then select DONE.

15

PACING:

EVENT PACED

3 EVENT CONDITIONS

Proceed with the rest of the programming steps.