1 setting the clock and calendar, 2 menu screens: clock start times – Teledyne 6712SR User Manual

Page 48

Advertising
background image

6712SR Refrigerated Sampler
Section 3 Getting Started

3-6

When setting the clock and calendar, use a 24-hour clock for
times and the day-month-year format for dates. The illustration
at the left shows blanks for each position that accepts an entry.
The positions are called fields. When this screen appears on your
sampler, the fields will contain the sampler’s current settings for
the time and date.

• To move from one field to another without changing the

setting, press the arrow keys. Use this technique when
you want to change only one or two settings.

• Change the setting by typing a new number. Press

↵ to

accept the new setting.

For example, to enter 2:00 p.m. (2:00 p.m is 14:00 on a 24-hour
clock), type 14. Press

↵. Next, type 0 (zero) for the minutes, and

press

↵. To enter a date, such as January 22, 2001, type: 22 ↵

01

↵ 01 ↵ .

3.4.1 Setting the Clock and

Calendar

3.4.2 Menu Screens: Clock

Start Times

There are four start time settings you can use in a sampling
program. This section discusses only one of these settings, the
clock start time. A clock start time lets you set up a sampling
program that starts at the same time on one or more days of the
week. For example, you may want to begin sampling at 6:00 a.m.
on Monday through Friday (refer to the screen below).

To select a day or days from the menu:

1. Press an arrow key until the cursor is on the correct day.

Then, press

↵.

2. Repeat these steps until each day you want is blinking.

3. Press an arrow until DONE blinks. Press

↵.

Entering Times and Dates

ENTER TIME AND DATE:

HH:MM DD-MON-YY

__:__ __-__-__

•Move from one field to

another by pressing the
arrow keys.

•Change a setting in a field

by typing a number. Press

to accept the new setting.

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Select OTHER FUNCTIONS.

2

MAINTENANCE

MANUAL FUNCTIONS

PROGRAMMING STYLE

Select MAINTENANCE.

3

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Select SET CLOCK.

4

ENTER TIME AND DATE:

HH:MM DD-MON-YY

14:00 22-JAN-01

Enter the time and date.

5

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Press Stop to return to the main
menu.