6712SR Refrigerated Sampler
Section 5 Extended Programming

5-8

23

0 RINSE CYCLES

(0-3)

Enter 0.

24

RETRY UP TO 0 TIMES

WHEN SAMPLING

(0-3)

Enter 0.

25

ONE-PART PROGRAM

TWO-PART PROGRAM

Select TWO-PART PROGRAM.

26

24 BOTTLES AVAILABLE

ASSIGN BOTTLES

1 THRU 6 TO

PART ‘A’ (1-23)

Enter 6.

Screen will say "Beginning Part A" before proceeding with screen 27.

27

UNIFORM TIME PACED

FLOW PACED

EVENT PACED

NONUNIFORM TIME

Select UNIFORM TIME PACED.

28

TIME BETWEEN

SAMPLE EVENTS:

0 HOURS, 5 MINUTES

Enter 0 for hours, 5 for minutes.

29

1 BOTTLES PER

SAMPLE EVENT (1- 6)

Enter 1.

30

SWITCH BOTTLES ON:

NUMBER OF SAMPLES

TIME

Select NUMBER OF SAMPLES.

31

SWITCH BOTTLES EVERY

1 SAMPLES (1-50)

Enter 1.

32

RUN CONTINUOUSLY?

YES

NO

Select NO.

33

DO YOU WANT

SAMPLE VOLUMES

DEPENDENT ON FLOW?

YES

NO

Select NO.

34

SAMPLE VOLUME

200 ml (10-1000)

Enter 800.

Table 5-2 Storm Water Runoff Sampling (Continued)