Section 7 Remote Operation

7-5

23 = VOLUME OUT OF RANGE. (the sample volume

requested is outside its range (10-9990 ml)

STI,35523.41875, is the most recent sample time.

BTL,2, is the bottle receiving the most recent sample.

SVO,100, is the most recent sample’s volume.

SOR,0, is the results of attempting the most recent sample.
The possible results are:

· 0 = SAMPLE OK.

· 1 = NO LIQUID FOUND.

· 2 = LIQUID LOST (not enough liquid to satisfy

requirements.)

· 3 = USER STOPPED (using the Stop Key).

· 4 = POWER FAILED.

· 5 = PUMP JAMMED.

· 6 = DISTRIBUTOR JAMMED.

· 8 = PUMP LATCH OPEN.

· 9 = SAMPLER SHUT OFF (while sampling).

· 11 = NO DISTRIBUTOR.

· 12 = SAMPLE IN PROGRESS.

CS,xxxx is the byte-by-byte sum of the entire string
“MO,...,CS,”

Note

For commands that are not understood (INVALID COMMAND
or CHECKSUM MISMATCH) or when no samples have been
taken, the most recent bottle information is left off the response
string.

Send Data – When receiving this command, the sampler
returns the Send Status appended with data from any module,
rain gauge, or SDI-12 device. Refer to Table 7-2 to interpret the
returned string.

Table 7-2 Serial Data Codes

Identifier Parameter

Units

B?

1

Bottle number and time

days since 1-Jan-1900

CR

Current day’s rain

tips

2

CS

3

Checksum

unsigned long

DE

Description

6712 Sampler

FL

Flow

cubic meters per second

ID

Unit specific identifier

10 digit unsigned long

LE

Level

meters

MO

Model

6712