Section 5 Extended Programming

5-11

The above example presented several screens that are not
available in standard programming. Descriptions of features
available in extended programming follow.

61

RESET RAIN HISTORY?

YES

NO

Select YES.

62

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

Select SET POINT.

63

“LEVEL” SET POINT

0.25 ft

(0.100-30.000)

Enter 0.25.

64

ONCE ENABLED,
STAY ENABLED?

YES NO

Select YES.

65

SAMPLE AT ENABLE?

YES NO

Select NO.

66

START OF SAMPLING

AFTER A

15 MINUTE DELAY

(1-999)

Enter 20.

67

PAUSE RESUME

1. HH:MM DD HH:MM DD
2. HH:MM DD HH:MM DD

CLEAR DONE

Select DONE.

Screen will say “End of Part B.”

68

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

Select DELAYED START.

69

START “STORM”

AFTER A 72 HOUR

“DRY” PERIOD?

YES NO

Select NO. (Option for samplers using a 0.01 inch tip rain gauge. See Dry
Period Option, described on page 5-26.)

70

START “STORM”

AFTER A

1 MINUTE DELAY

(1-999)

Enter 1.

71

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Select NO.