Teledyne 6712SR User Manual

Page 85

Advertising
background image

6712SR Refrigerated Sampler

Section 4 Standard Programming

4-31

8

MODE OF OPERATION:

FLOWMETER

LEVEL ONLY

If you choose Flowmeter, proceed to screen #9 below.
If you choose LEVEL ONLY, proceed to screen #12.

9

WEIR FLUME

DATA POINTS

EQUATION MANNING

FLOW-INSERT

When you select Flowmeter, you will have several options for performing
the flow conversion.
For this example, choose Weir. Press

.

10

V-NOTCH

RECTANGULAR

CIPOLLETTI

For this example, choose V-Notch. Press

.

11

V-NOTCH WEIR ANGLE:

120 90 60

45 30 22.5

For this example, choose a Weir angle of 90. Press

.

12

CURRENT LEVEL IS

__.__ ft

ADJUST LEVEL TO

__.__ ft

The current level will be displayed. Enter an adjustment if needed. If the
adjustment is different by more than 0.50 ft, you will be prompted “

ARE

YOU

SURE

?

YES

NO

” Choose

NO

to re-enter; select

YES

if correct. Press

.

13

NUMBER OF BOTTLES:

1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press

. For this example, select 24.

14

BOTTLE VOLUME IS

1000 ml (300-30000)

Type the volume for the bottles in your kit. For this example, 1000 is
correct, so simply press

.

15

SUCTION LINE LENGTH

IS 10 ft

(3-99)

Type the length of the suction line, then press

.

If you change the length, the sampler will display a message,
“PLEASE WAIT!. . . GENERATING PUMP TABLES.”

16

TIME PACED

FLOW PACED

Select FLOW PACED.
If the module is being used as a flowmeter, you will see screen #17a.
If the module is being used for level only, you will see screen #17b.

17

a

FLOW BETWEEN

SAMPLE EVENTS:

0.075 Mgal

(0.001 - 99999)

If the module is being used as a flowmeter, choose the amount of flow
between sample events.

Table 4-4 Programming Example With 730 Module Installed (Continued)