1 sampler enable, 1 sampler enable -24, 4 sampler enable -24 – Teledyne 6712SR User Manual

Page 112: Table 5-4 sampler enable

Advertising
background image

6712SR Refrigerated Sampler
Section 5 Extended Programming

5-24

5.12.1 Sampler Enable

This example uses a sampler with an attached 730 Bubbler Flow
Module, Rain Gauge and an SDI-12 sonde at address 5 with four param-
eters.

Table 5-4 Sampler Enable

1

ENABLE:

NONE PROGRAMMED

In this example, you will be enabling the sampler if it rains .5” in 2 hours
and the level goes above 1 foot.

2

ENABLE:

RAIN LEVEL FLOW

TEMP5 pH5 SP_CO5

DO5 NONE

Select RAIN.

3

ENABLE: RAIN

AND OR DONE

Choose AND.

4

ENABLE: RAIN AND

LEVEL FLOW TEMP5

pH5 SP_CO5

DO5 DONE

Select LEVEL.

5

“RAIN” SET POINT:

.5 INCHES PER

__HOURS, __MINUTES

Enter .5 for the inches of rain. Press

.
Enter 2 HOURS.

6

ENABLED WHEN:

ABOVE SET POINT
BELOW SET POINT

Select ABOVE SET POINT.

7

RESET RAIN HISTORY?

YES NO

Select YES.

8

“LEVEL” CONDITION

SET POINT

RANGE

RATE OF CHANGE

Select SET POINT.

9

“LEVEL” SET POINT:

+1 ft

( 0.001 - 30.00)

Enter 1 foot.

10

ENABLED WHEN:

ABOVE SET POINT
BELOW SET POINT

Select ABOVE SET POINT.

11

ENABLE:

RAIN >0.50” / 2:00

AND

LEVEL >1.000 ft

Your enable conditions will be summarized on the display.
If correct, press

to continue.

If necessary, make changes to the enable conditions.