Section 4 Standard Programming

4-3

The following table is an example of flow-paced sampling.

• Program type: Standard

• Site description: SITE 29

• Bottle kit: 24, 1000 milliliter bottles

• Suction line length: 7 feet

• Pacing: Flow pacing, every two pulses

• Distribution: 2 bottles per sample

• Sample Volume: 250 milliliters

• Start time: Clock Time, 6:00 a.m. on Monday,

Wednesday, Friday

• Module: No module installed

11

SAMPLE VOLUME:

200 ml (10-1000)

Type the volume of the sample you want deposited in each bottle. Then,
press

.

12

NO DELAY TO START

DELAYED START

CLOCK TIME

WAIT FOR PHONE CALL

For this example, select DELAYED START by pressing an arrow until the
option blinks. Then, press

.

13

FIRST SAMPLE

AFTER A

5 MINUTE DELAY

(1-999)

Type the delay period you want between the time you run the program and
the time the sampler takes the first sample. Then, press

.

14

PROGRAMMING COMPLETE

RUN THIS PROGRAM

NOW?

YES NO

Run the program immediately by selecting YES. Select NO if you want to
run the program later by selecting RUN from the main menu. Press

after

making your choice. In this example, NO is selected.

15

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Run the program by selecting RUN and pressing

.

Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle (Continued)

Table 4-2 Standard Program: Flow-Paced Sampling, Two Bottles Per Sample

1

6712 SAMPLER

STANDARD PROGRAMMING

For HELP at any

screen press ? key.

2

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Select PROGRAM.

3

SITE DESCRIPTION:

“FACTORY051“

CHANGE?

YES NO

Select NO.