Section 4 Standard Programming

4-17

4.13 Operating the Pump

You can manually operate the sampler’s pump in the forward or
reverse direction. You can do this to test the pump, draw a large
sample, or purge the suction line.

When you run the pump manually, the sampler reports the
number of pump counts during the pump’s operation. Use this
feature to obtain an estimate of the pump counts required to
purge a nonstandard suction line.

You can also manually run the pump from the main menu. Press
the “1” key and then

↵ to run the pump in reverse. Press the “3”

key and then

↵ to run the pump forward.

Calibrating Sample Volumes

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

•Select OTHER FUNCTIONS

2

MAINTENANCE

MANUAL FUNCTIONS

MODULE

HARDWARE

•Select MANUAL FUNCTIONS

3

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

•Select CALIBRATE VOLUME

4

SAMPLE VOLUME:

___ ml (100-9990)

•Enter the desired volume

5

CALIBRATE VOLUME

PRESS

↵ WHEN READY!

6

PURGING

SUCTION LINE

7

PUMPING 200 ml

8

PURGING

SUCTION LINE

9

VOLUME DELIVERED:

____ml

•Measure the volume delivered

and enter the amount here.

10

200 ml

ARE YOU SURE?

YES NO

This screen appears when the
amount you entered in step 9
was more than twice, or less
than half, the programmed
volume.

• Select YES when the volume

delivered matches the volume
displayed on this screen.

• Select NO to enter the volume

delivered.