4 maintenance screens, 1 set clock, 4 maintenance screens -4 – Teledyne 6712SR User Manual

Page 152: 1 set clock -4, 1 setting the clock and calendar -4, Table 8-1 setting the clock and calendar

Advertising
background image

6712SR Refrigerated Sampler
Section 8 Maintenance

8-4

8.4 Maintenance Screens

When you select OTHER FUNCTIONS from the main menu, and
then select Maintenance, the following screen will be displayed:

Using these options, you can set the sampler’s time and date, set
the number of pump counts for the pump tube warning, reset the
battery warning for the internal battery, and run a set of diag-
nostic tests for troubleshooting purposes.

8.4.1 Set Clock

Teledyne Isco ships samplers from the factory with the clocks set
to the correct Central Standard Time. Reset the clock when
installing the sampler in a different time zone, for daylight
savings time, or when needed.

When setting the clock and calendar, use a 24-hour clock for
times and the day-month-year format for dates. The illustration
at the left shows blanks for each position that accepts an entry.
The positions are called fields. When this screen appears on your
sampler, the fields will contain the sampler’s current settings for
the time and date.

• To move from one field to another without changing the

setting, press the arrow keys. Use this technique when
you want to change only one or two settings.

• Change the setting by typing a new number. Press

↵ to

accept the new setting.

For example, to enter 2:00 p.m. (2:00 p.m is 14:00 on a 24-hour
clock), type 14. Press

↵. Next, type 0 (zero) for the minutes, and

press

↵. To enter a date, such as January 22, 2001, type: 22 ↵

01

↵ 01 ↵ .

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Entering Times and Dates

ENTER TIME AND DATE:

HH:MM DD-MON-YY

__:__ __-__-__

•Move from one field to

another by pressing the
arrow keys.

•Change a setting in a field

by typing a number. Press

to accept the new setting.

Table 8-1 Setting the Clock and Calendar

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Select OTHER FUNCTIONS.

2

MAINTENANCE

MANUAL FUNCTIONS

PROGRAMMING STYLE

Select MAINTENANCE.

3

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Select SET CLOCK.

4

ENTER TIME AND DATE:

HH:MM DD-MON-YY

14:00 22-JAN-01

Enter the time and date.

Advertising