Section 4 Standard Programming

4-15

4.11 Grab Samples

Grab samples let you take a single sample on demand, collecting
the sample in an external container (Figure 4-4). Keep in mind:

When the sampler delivers a grab sample, it runs through a com-
plete sampling cycle, using the current settings for volume and
for line rinses or retries. If it is a two-part program, the sampler
uses the settings for part A.

To take a grab sample while the sampler is running a program,
you must interrupt the program by pressing the Stop key.
Restart the program by running it again.

To take a grab sample:

1. Disconnect the pump tube from the bulkhead fitting.

2. Place the end of the tube over a collection container.

3. Follow the steps in Taking a Grab Sample at left.

4. Reconnect the pump tube to the bulkhead fitting.

Figure 4-4 Taking a Grab Sample

Taking a Grab Sample

1

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

•Select GRAB SAMPLE

2

SAMPLE VOLUME:

___ ml (10-9990)

•Enter the desired volume

3

GRAB SAMPLE

PRESS

↵ WHEN READY!

4

PURGING

SUCTION LINE

5

PUMPING 200 ml

6

PURGING

SUCTION LINE

7

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

•Press STOP to return to

main menu.