Section 6 SDI-12 Sondes

6-2

Figure 6-1 SDI-12 Sonde Parameters

6.2 Scanning For Sondes

You must be in Extended Programming to use a sonde with the
6712. The sampler can automatically scan for the presence of
sondes if you enable that option:

• From the main menu, select OTHER FUNCTIONS.

• Select HARDWARE.

• As you press

↵ to move through the hardware screens,

you will see one that looks like the following:

Select “NO SDI-12 SONDE” to scan for sondes. If one is found,
you can set it up, and the “AUTO SDI-12 SCAN” option will be
switched to yes.

Select the “AUTO SDI-12 SCAN” option to manually turn on (or
off) automatic scanning for sondes when the sampler is turned
on.

If a sonde has already been connected and set up, the hardware
screen will instead display a screen similar to the following:

SDI-12 Sonde Data Units:

Parameter

Range

Resolution

Storage Bytes

Temperature

-40 - 100 °C

0.1 °C

2

Conductivity

0 - 100 mS/cm

0.01 mS/cm

2

Specific Conductance

0 - 100 mS/cm

0.01 mS/cm

2

Total Dissolved Solids

0 - 90 g/l

0.01 g/l

2

Salinity

0 - 70 ppt

0.1 ppt

2

Dissolved Oxygen

0 - 20 mg/l

0.1 mg/l

1

pH

0 - 14

0.1

1

ORP

-999 - 999 mv

0.1 mv

2

Level

0 - 6.5279 m

0.0001 m

2

Ammonium-Nitrogen

0 - 200 mgN/l

0.1 mgN/l

2

Ammonia-Nitrogen

0 - 200 mgN/l

0.1 mgN/l

2

Nitrate-Nitrogen

0 - 200mgN/l

0.1 mgN/l

2

Turbidity

0 - 5000 NTU

0.1 NTU

2

Chloride

0 - 1000 mg/l

0.1 mg/l

2

Chlorophyll

0-400 ug/l

0.1 ug/l

2

Other

-99999.0-99999.0 0.0001

4

NO SDI-12 SONDE

AUTO SDI-12 SCAN OFF

SDI-12 DATA:

TEMP0 pH0 SD_CO0

SAL0 05DATA0 DO0

ORP0 LEVEL0 09DATA0