6712SR Refrigerated Sampler

Section 4 Standard Programming

4-33

Table 4-5 Programming Example With 750 Module Installed

1

MODULE INSERTED---

DOWNLOAD DATA NOW

OR LOSE ALL DATA!

DONE

With the sampler off, insert the module in the bay on the controller.
Turn the sampler on by pressing the

On/Off

key. Press

. Download stored

data, if you want to keep it. Pressing DONE erases it and advances to the
next screen.

2

6712 SAMPLER

STANDARD PROGRAMMING

For HELP at any

screen press ? key.

This screen will now appear for 8 seconds.

3

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

The option PROGRAM will be blinking. Press

.

4

SITE DESCRIPTION:

“FACTORY051“

CHANGE?

YES NO

The option NO will be blinking. Press

.

For the purposes of this example, it isn’t necessary to change the description.

5

SELECT UNITS FOR

FLOW RATE:

cfs gps gpm Mgd

lps m3s m3h m3d

Select the desired units for the flow rate. For this example, choose cfs.
Press

.

6

SELECT UNITS FOR

FLOW VOLUME:

cf gal Mgal

m3 lit

Select the desired units for the flow volume. For this example, choose Mgal.
Press

.

7

SELECT UNITS FOR

VELOCITY:
fps mps

Select the desired units for the flow velocity. For this example, choose fps.
Press

.

8

PROGRAM MODULE?

YES NO

Select YES to program the module.

9

MODE OF OPERATION:

FLOWMETER

LEVEL AND VELOCITY

The 750 can have two different modes of operation: Flowmeter or Level and
Velocity.
For this example, choose Flowmeter. Press

.

10

FLOW CALCULATION:

AREA * VELOCITY

LEVEL TO FLOW RATE

When you select Flowmeter, you will have two options for performing the
flow conversion: Area * Velocity or Level to Flow Rate.
For this example, choose Area * Velocity. Press

.

11

LEVEL TO AREA:

CHANNEL SHAPE

DATA POINTS

For this example, choose Channel Shape. Press

.

12

ROUND PIPE

U-CHANNEL

RECTANGULAR CHANNEL

TRAPEZOIDAL CHANNEL

For the channel shape, select Round Pipe. Press

.