background image

6712SR Refrigerated Sampler

Section 2 Installation/Preparation

2-5

Figure 2-2 Installing the Bottle Racks

Figure 2-3 Installing the 24-bottle, 12-bottle, and 8-bottle Kits

24-Bottle Kit

1000-milliliter polypropylene bottles.

2-Bottle Kit

(Requires a locating base.)

Stop

(1 of 5)

Ramps

Install the distributor shaft extension.
Use the 8

11

/

16

-inch extension for the 12-

and 24-bottle kits. Use the 10

1

/

2

-inch ex-

tension for the 8-bottle kit. The instruc-
tions on page 2-1 show you how to remove
the controller and replace the extensions
for each bottle kit.

Next, adjust the distributor arm to position
D for the 12-bottle and 24-bottle kits, or to
position B for the 8-bottle kit. Install the
arm and stop. See Installing the Distribu-
tor Arm and Discharge Tube on page 2-2.

Slide the rack, with bottles installed onto
the mounting plate. Adjust the stops and
ramps on the plate so that the rack is cen-
tered and held firmly on the plate.

Finally, position the arm over each bottle.
If the arm does not stop over each bottle,
readjust the rack.

NEVER MANUALLY ROTATE

THE DISTRIBUTOR ARM.