6712SR Refrigerated Sampler
Section 4 Standard Programming

4-30

CAUTION

Most Teledyne Isco accessories and interfacing equipment are
not certified for us in hazardous locations as defined by the
National Electrical Code. Check applicable guidelines, codes,
and regulations of federal, state, city, and county agencies.
Refer to Appendix C in the back of this manual for more safety
information.

Table 4-4 Programming Example With 730 Module Installed

1

MODULE INSERTED---

DOWNLOAD DATA NOW

OR LOSE ALL DATA!

DONE

With the sampler off, insert the module in the bay on the controller.
Turn the sampler on by pressing the

On/Off

key. Press

.

Download stored data, if you want to keep it. Pressing DONE erases it and
advances to the next screen.

2

6712 SAMPLER

STANDARD PROGRAMMING

For HELP at any

screen press ? key.

This screen will now appear for 8 seconds.

3

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

The option PROGRAM will be blinking. Press

.

4

SITE DESCRIPTION:

“FACTORY051“

CHANGE?

YES NO

The option NO will be blinking. Press

.

For the purposes of this example, it isn’t necessary to change the
description.

5

SELECT UNITS FOR

FLOW RATE:

cfs gps gpm Mgd

lps m3s m3h m3d

Select the desired units for the flow rate. For this example, choose cfs.
Press

.

6

SELECT UNITS FOR

FLOW VOLUME:

cf gal Mgal

m3 lit

Select the desired units for the flow volume. For this example, choose Mgal.
Press

.

7

PROGRAM MODULE?

YES NO

Select YES to program the module.