VEGAPULS 42, 44 and 45 – VBUS

77

Adjustment with the PC on VEGALOG

If VVO (adjustment software) gets in contact
with the processing system for the first time,
you will be asked if you want to transfer the
data from the processing system to the PC.

• Click to "

Yes".

If you have connected the VVO to a
VEGALOG from which data has already been
saved, you are asked if the saved data
should be transferred from the PC to
VEGALOG or the data from VEGALOG to the
PC.

Configuration

The adjustment of a radar sensor with the PC
on a VEGALOG generally corresponds to the
adjustment of a radar sensor on the signal
conditioning instrument, as in chapter "5.2
Adjustment with the PC on VEGAMET".

The only difference are the increased con-
figuration requirements of VEGALOG. A
number of possible sensor inputs and signal
outputs as well as different processing
routines must be assigned. First of all, you
must define where and what processing
which sensor (input) does.

Configuration info

• Choose the menu "

Configuration/Measur-

ing system".

The preadjusted user identification can be
modified at a later time in the menu "

Configu-

ration/Program/User access".

No matter whether you set up a measuring
system (unit consisting of sensor and
VEGAMET signal conditioning instrument or
VEGALOG processing system) with the
adjustment software VVO or with the signal
conditioning instrument, the adjustment pro-
cedure is always the same:
- first of all, create a measurement loop and

configure a measuring system in the menu
"

Configuration" and then

- carry out the parameter adjustment of the

sensor in the menu "

Instrument data".

You now reach the menu "

Configuration

Measuring system", in which a graphical
representation of the VEGALOG connected
to the PC is displayed.

• Click to an individual card number and you

will get a "Card info" in the window

.