Set-up, configuration – VEGA VEGAMET 514… D User Manual

Page 22

Advertising
background image

22

VEGAMET 514… D

Set-up, configuration

Monitoring

For the activation of the monitoring the following pre-
requirements must be met:
- Determine in the menu range "Configuration

meas. point" input 4

- Connect the key switch (opener) to input 4 (see

page 13).

- Select in the menu range "Parameter adjustment

relay outputs" the monitoring mode for the
concerned relay (see page 37 and 40).

The selected relay is energized.

By pushing the key switch (opener opens) the respectively actual measured value is frozen and derived from
this, in case the value is decreased (low) or exceeded (high) the relay energizes.

After a reset and pushing the key switch again the actual measured value is frozen again etc.

–– MST1

–– Monitoring

–– Input

not defined

Input 4

with input 4 determination of the correction signal

input (key switch)

–– MST2

as described above

Tare

By means of the tare function an additional
measured value exists from the actual value
beginning at 0%. The tare function can be triggered
via a key connected to correction signal input 4.

Each current, voltage or relay output as well as the
display can be afterwards coordinated to the tared
measured value in the menu point "relating to". With
the allocation or allocations the tared measured
value is activated.

–– MST1

–– Tare

–– Input

not defined

Input 4

Determination of the correction signal input (key)

Meas.value

50 %

High 7 %
Low –5 %

>57 %

Relay de-

energized

<45%

Relay de-
energized

Relay energized

100 %

0 40 100 %

0

40

Measured values in %

Key

tared course

actual course

tared measured values

actual measured values

Advertising