VEGAMET 514… D

Product description

Current output

Number

3 outputs

Function

current output 2 and 3 can be configured and
parameter adjusted as:
- level proportional
- flow proportional

- flow pulses
- sampling pulse

Output values
- level / flow proportional

Load

max. 500

Range

0/4 … 20 mA

Resolution

0,05 % of range

Linearity error

0,05 % of range

Temperature error

0,05 %/10 k of range

- pulse output

Load

min. 600

Voltage pulse

12 V DC at 20 mA

Pulse length

200 ms

Voltage output

Number

3 outputs

Function

analog output of the output results

Range

0/2 … 10 V

Current

max. 1 mA

Resolution

0,05 % of range

Linearity error

0,05 % of range

Temperature error

0,06 %/10 k of range

Relay outputs

Number

2 switching relays
1 fail dafe relay

Function

can be configured and parameter adjusted as:
- switching relay for level or flow
- pulse relay-flow quantity
- pulse relay-sampling

Contact

1 floating spdt

AgNiO and hard gold plated

Turn-on voltage

min. 10 mV DC
max. 250 V AC/DC

Switching current

min. 10

µ

A

max. 3 A AC, 1 A DC

Breaking capacity

max. 500 VA, 54 W

Min. switching hysteresis (Low/High-Delta)

0,5 %

DISBUS-output

Function

digital transmission of output results and

system information to VEGADIS 174 or
VEGACOM 557

Connection line

2-wire

Line length

max. 1000 m

>>