8 Application examples

VEGAMET 391 • 4 … 20 mA/HART

36032-EN-130701

Select in the DTM window "Linearization" the option "Flow" and then

the used flume (in the above example V-notch).

Select in the DTM window "Scaling" under "Parameter" the option

"Flow". Finally the allocation of a value must be carried out, i.e. the

flow volume is assigned to the 0 and 100 % value. As the last step,

select the requested meas. unit. For above example: 0 % = 0 and

100 % = 400, meas. unit m³/h.

First of all decide if you want to use a relay and/or a current output. In

the DTM window "Outputs" you can use any of the three outputs as

long as these are not yet used for other tasks.
Finally select under "Mode" (relay) or "Output characteristics" (cur-

rent output) the option "Flow volume pulse" or "Sampling pulse". Enter

under "Pulse output all" the flow volume after which a pulse should be

outputted (e.g. 400 m³ corresponds to one pulse per hour with a flow

volume of 400 m³/h).
In mode "Sampling pulse" an additional pulse is outputted after a

defined time. This means that a timer is started after each pulse after

which another pulse is outputted. This applies only if already a pulse

was outputted after exceeding the flow volume.
Due to sludge at the bottom of the flume, it can happen that the min.

adjustment originally carried out can no longer be reached. Therefore

small quantities will continuously enter the flow volume detection

despite the "empty" flume. The option "Min. flow volume suppression"

offers the possibility to suppress measured flow volumes below a

certain percentage value for the flow volume detection.

If a flow measurement is set up, the flow value can also be summed

up and displayed as flow volume. The graph can be selected in the

menu item "Display". The following parameters must be adjusted for

the totalizer:

Measuring unit: Selection of the unit by which the totalizer adds.

Indicating format: Selection of the indicating format (number of

decimal positions of the counter)

Information:

The totalizer can be reset in the menu "Additional settings" - "Reset"

Linearization curve

Scaling

Outputs

Totalizer