Required laboratory reagents, Luminex is 2.3 software, Luminex 200 performance specifications – Luminex 200 User Manual with IS 2.3 Software User Manual

Page 19: Speed -3 accuracy and precision -3

Advertising
background image

x

MAP Technology

The System

PN 89-00002-00-144 Rev. B

2 - 3

Required

Laboratory

Reagents

Household bleach

70% isopropanol or 70% ethanol

Mild detergent

Distilled water

Luminex IS 2.3

Software

Luminex IS 2.3 software provides complete control of the system
and performs data analysis. Your Luminex 200 system is preloaded
with the Luminex software. However, we supply a software CD
should you need to reinstall the software.

This software requires a dedicated system. Unauthorized additional
software is prohibited and may result in improper operation of the
system.

Luminex 200

Performance

Specifications

Speed

Speed: greater than or equal to1.7 GHz Intel® Pentium® IV
processor with 256 MB RAM

USB communications link for fast data transfer

Automatic transfer of assay templates and new reagent
information into the system via a large capacity read/write CD

Installation: < 4 hours

System calibration: < 10 minutes

System controls: < 10 minutes

Barcode reader entry of sample IDs

Automatic post-analysis

Analyze one 96-well plate/hour depending on manufacturer’s kit

Up to 100 xMAP microsphere sets per sample

System warmup: 30 minutes. Systems that remain inactive for at
least four hours will require a warm-up to restart the lasers. After
acquiring sample, running system calibrators, running system
controls, and warming up the instrument, the system resets the
four-hour internal clock.

Accuracy and Precision

Sample uptake volume: ± 5%

Classification of xMAP microspheres: > 80%

Misclassification of xMAP microspheres:

≤ 2% - may vary by

xMAP microsphere product lines. Refer to the specific product
information sheet for further details.

Temperature control: 0°C to + 2°C of target