Luminex 200 System User Manual

x

MAP Technology

2 - 10

PN 89-00002-00-144 Rev. B

Air Intake Filter

A replaceable air intake filter cleans the air used to pressurize sheath
fluid. This filter is enclosed behind an access door located on the
back of the Luminex 200 analyzer.

Syringe

The syringe delivers a sample from the 96-well microtiter plate to the
cuvette.

Sheath Filter

The sheath filter removes particles greater than ten microns in
diameter from the sheath fluid.

Air, Waste, and

Sheath Fluid

Connectors

The air, waste, and sheath connectors, located on the left side of the
analyzer, connect to the SD system and waste fluid containers using
clear tubing. The air connector is green, the sheath fluid connector is
blue, and the waste fluid connector is orange.

Luminex Sheath

Delivery System

For proper operation, place the Luminex SD system at the same level
as the base of the Luminex XYP instrument. Do not put it on top of
the Luminex 200 analyzer. If you are not using the SD system, sheath
fluid levels must be monitored manually. Check the sheath fluid level
before starting a run or procedure.

Waste Fluid

Container

The waste fluid container receives waste from the system. The waste
container should not be placed on top of the instrument. Ensure that
the waste tubing is not elevated above the level of the Luminex 200
analyzer at any point. Contact Luminex Technical Support before
you relocate the waste fluid container. To maintain a stable flow rate,
do not move the waste line or container during system operation.

Optical

The optical system consists of the optical assembly and the excitation
lasers. The optical assemblies do not require manual adjustment by
the user.

xMAP Technology
Reagents

The xMAP technology reagent system consists of classification
calibration microspheres, reporter calibration microspheres,
classification control microspheres, and reporter control
microspheres.

Warning:

If biological samples have been tested with the

system, use your standard laboratory safety practices.

Caution:

Waste levels must be manually monitored. Do not

allow the waste container to overflow

.