Malfunction -17 draining the reservoir -17, Malfunction, Draining the reservoir – Luminex 200 User Manual with IS 2.3 Software User Manual

Page 59

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-144 Rev. B

4 - 17

Malfunction

If the alarm sounds even though the bulk sheath container has fluid
and the sheath filter is in good condition, the system is reporting a
malfunction. If this happens, contact Luminex Technical Support.

Draining the Reservoir

If you need to ship the Luminex SD system back to Luminex
Corporation, drain the reservoir before you pack the system.

1. In the IS 2.3 software, click Prime or Warmup to pressurize the

system.

2. Leave the green air tubing connected between the Luminex 200

analyzer and the SD system.

3. On the SD front panel, disconnect the blue tubing from the

intake labeled Sheath Out and disconnect the white tubing from
the intake labeled Sheath In.

4. Insert the white tubing into the intake labeled Sheath Out and

insert the blue tubing into the intake labeled Sheath In.

5. Turn the unit off and then on again.

6. Press the Prime button on the front panel of the SD system.

7. Sheath fluid will pump from the Sheath Delivery System

reservoir into the 20L Sheath Box.

To refill the Sheath Delivery System, reconnect the tubing according
to color coding and press the Prime button on the SD.

Contact Technical Support for additional information.