Troubleshooting

PN 89-00002-00-144 Rev. B

4 - 13

Low Voltage Detected

Possible laser failure.

Turn off the analyzer, XYP, and
PC, then turn them back on.
Make sure system is completely
warmed up. Calibrate and verify
the system. If the error message
occurs again, contact Luminex
Technical Support.

Runtime Sheath Pressure out
of Limits (Too High)

The sheath fluid pressure is too
high.

Ensure that the sheath fluid
container is at the same level as
the analyzer. On systems with a
Luminex Sheath Delivery
system, ensure that the sheath
fluid pressure reading is equal to
the original reading. If it is not,
contact Technical Support for
instructions to adjust the
regulator. Calibrate and verify
the system.

Runtime Sheath Pressure Out
of Limits (Too Low)

There is a pressurization
problem with the sheath fluid.

See Pressurization on page 4-3.

Low Laser Power Detected

The system is not sufficiently
warmed up.

Warm up the system. This takes
approximately 30 minutes.
Calibrate and verify the system.

Possible Laser Failure

Open the calibration trend
report for CAL2 and record the
last three reporter voltages.
Contact Luminex Technical
Support with this information.

Command Encountered Time
Out Error

The current command failed.

Check the system to verify that
it is warmed up and that there
are no pressurization problems.
Repeat the command. If the
command continues to time out,
contact Technical Support.

Instrument Not Calibrated

The instrument is out of
calibration.

Calibrate the instrument.