Luminex xyp instrument general, Optics -5 fluidics -5 electronics -5, Luminex xyp instrument general -5 – Luminex 200 User Manual with IS 2.3 Software User Manual

Page 21: Optics, Fluidics, Electronics

Advertising
background image

x

MAP Technology

The System

PN 89-00002-00-144 Rev. B

2 - 5

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively

Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240
V~ ± 10%, 0.8 Amp, 47-63 Hz.

ac inlet fuse: 3 Amp, 250 V~, fast acting

Optics

Reporter laser: 532 nm, nominal output 10-16.5 mW, maximum
500 mW, frequency-doubled diode; mode of operation,
continuous wave (CW)

Classification laser: 635 nm, 9.1 mW ± 6%, maximum output 25
mW, diode; mode of operation, continuous wave (CW)

Reporter detector: Photomultiplier tube, detection bandwidth of
565-585 nm

Classification detector: Avalanche photo diodes with temperature
compensation

Doublet discrimination detector: Avalanche photo diodes with
temperature compensation

Fluidics

Sheath flow rate 90 µL ± 5 µL/second

Cuvette: 200 micron square flow channel

Sample injection rate: 1 µL/second ± 0.05 µL

Sample uptake volume: 20-200 µL

Electronics

Reporter channel detection: A/D resolution 14 bits

Communications interface: USB

Luminex XYP instrument, communications interface: RS 232

Luminex XYP

Instrument General

Ambient temperature: 15°C to 30°C (59°F to 86°F)

Humidity: 20% to 80%, noncondensing

Altitude: operation up to 2400 m (7874 ft) above mean sea level

Physical dimensions: 44 cm (17.25 inches) W x 60 cm (23.5
inches) D x 8 cm (3 inches) H

Weight: 15 kg (33 lbs.)

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

Heater operating range: 35°C to 55°C (95°F to 131°F) with
tolerance 0°C to +2°C